TrackSense Pro Basic draadloze datalogger - dubbele starre/flexibele temperatuur sensor

TrackSense Pro Basic draadloze datalogger - dubbele starre/flexibele temperatuur sensor

The wireless TrackSense® Pro data loggers consist of resistant stainless steel and offers high-end technology, which provides highly accurate measurements within different thermal processes.

The Double Rigid/Flex Temperature Sensor is equipped with both a smart flex sensor for hard-to-reach positions and a stainless steel bent-sensor (either bent 90  or 180 °). It has a temperature range of -30 °C to +105 °C with an unmatched accuracy of ±0.05 °C (in range -25 °C to +105 °C). It is ideal for applications within the food industry, such as heat penetration and heat distribution studies. 

Ellab’s TrackSense® Pro data loggers can all be configured with interchangeable sensors to measure pressure, relative humidity, temperature, CO2, and conductivity. The data loggers are easily activated and read by the Multi reader station and when combined with the Sky module they provide real time date via wireless communication. 

 

Contact Sales
Technical Specifications

Sensor:

Interchangeable Double Combi Temperature

Temperature Measuring Range:

-30 to +105 °C

Measuring Principle:

Electrical Resistance

Sensor Element:

Pt1000 

Diameter (Rigid / Flexible):

2 mm / 1.8 mm

Length:

From 10 mm / 100 mm

Position of Measuring Point from Round Tip:

3 mm

Accuracy:

-30 to -25 °C:

± 0.1 °C / ± 0.1 °C

-25 to +105 °C:

± 0.05 °C / ± 0.1 °C

Sensor Response Time:

T-63%:

0.8 / 2.2 Seconds

T-90%:

1.7 / 4.3 Seconds

Logger:

Pro Basic

Operating Temperature:

-30 to +105 °C

Operating Pressure*:

0.001 mBar to 4 Bar ABS

House Material:

316 L Stainless Steel

Diameter:

25 mm

Length:

44 mm

Weight With Battery:

48 Grams

Memory Capacity:

60,000 Data Points

Minimum Sample Rate:

1 Second

Maximum Sample Rate:

24 Hours

Maximum Start Delay:

14 Days

Intrinsically Safe:

Ex II1GD Ex ia IIC T3 Ga, -30°C ≤ Tamb ≤ +105°C

Time Accuracy:

± 5 Seconds Per 24 Hours

Battery:

TSP Standard Battery

If equipment is used in ATEX environment, special conditions for safe use stated in ATEX certificates section 17 must be considered.