jpgFeature/sdesk 696 implement raw brick http
TrackSense Pro XL Wireless Data Logger - Double Smart Flex Temperature Sensor

TrackSense Pro XL Wireless Data Logger - Double Smart Flex Temperature Sensor

The high-quality wireless TrackSense® Pro XL data loggers consist of resistant stainless steel and offers high-end technology that allows for stable and highly accurate measurements within different thermal processes. The XL logger is an enlarged logger with an extended battery capacity, which makes it exceptional for applications involving vast swings in temperature.

The Double Smart Flex Temperature Sensor has a temperature range of -196 °C to +140 °C with an accuracy of 0.1 °C (in range -40 °C to +140 °C). The sensor can be used within various industries for applications like environmental testing, and is ideal for situations involving hard-to-reach areas, as the Smart Flex sensors secure the intended position of the sensor.

All TrackSense® Pro data loggers are configured with interchangeable sensors able to measure temperature, relative humidity, pressure, conductivity and CO2. Ellab’s data loggers are easy to activate and read by the Multi reader station. When combined with the Sky module, data loggers can provide real time data via wireless communication.

Contact Sales
Technical Specifications

Sensor:

Interchangeable Double Smart Flex Temperature

Temperature Measuring Range:

-196 to +140 °C

Measuring Principle:

Electrical Resistance

Sensor Element:

Pt1000 

Diameter:

1.8 mm

Length:

100 - 1,000 mm (1,700 mm)

Position of Measuring Point from Round Tip:

3 mm

Accuracy:

-196 to -80 °C:

± 2.0 °C (± 0.5 °C if calibrated and adjusted at -196 °C)

-80 to -50 °C:

± 0.3 °C

-50 to -40 °C:

± 0.2 °C

-40 to +140 °C:

± 0.1 °C

Sensor Response Time:

T-63%:

2.2 Seconds

T-90%:

4.3 Seconds

Logger:

Pro XL

Operating Temperature:

-80 to +140 °C

Operating Pressure:

0.001 mBar to 4 Bar ABS

House Material:

316 L Stainless Steel

Diameter:

25 mm

Lenght:

68.8 mm

Weight With Battery:

80 Grams

Memory Capacity:

60,000 Data Points

Minimum Sample Rate:

1 Second

Maximum Sample Rate:

24 Hours

Maximum Start Delay:

14 Days

Intrinsically Safe:

Ex II1GD Ex ia IIC T3 Ga,  -55°C ≤ Tamb ≤ +105°C

Time Accuracy:

± 5 Seconds Per 24 Hours

Battery:

TSP 150L Battery

If equipment is used in ATEX environment, special conditions for safe use stated in ATEX certificates section 17 must be considered.