The wireless TrackSense® Pro XL data loggers consist of resistant stainless steel and offers high-end technology, which provides highly accurate measurements within different thermal processes. The XL logger is a larger logger with an extended battery capacity, which makes it ideal for applications involving vast swings in temperature.

The Double Flex High Temperature Sensor has a temperature range of 0 to +400 °C with an accuracy of 0.5 °C (in range 0 to +400 °C). The sensor can be used for various application, but is ideal for high temperature applications within the pharmaceutical industry, such as depyrogenation and ovens.

All of Ellab’s TrackSense® Pro data loggers are designed to accommodate for interchangeable sensors to measure temperature, conductivity, pressure, CO2, and relative humidity. The Ellab Sky module, when combined with a data logger, provides real time data via a wireless connection, while the Multi reader station allows users to easily activate and read data loggers. 

 

Contact Sales
Technical Specifications

Sensor:

Interchangeable Double Flex High Temperature

Temperature Measuring Range:

0 to +400 °C (>150 °C using Thermal Barrier)

Measuring Principle:

Electrical Resistance

Sensor Element:

Pt1000 

Diameter:

2 mm

Length:

150 - 1.000 mm

Position of Measuring Point from Round Tip:

3 mm

Accuracy:

0 to +400 °C:

± 0.5 °C

Sensor Response Time:

T-63%:

1.5 Second

T-90%:

2.9 Seconds

Logger:

Pro XL

Operating Temperature:

0 to +400 °C (>150 °C using Thermal Barrier)

Operating Pressure*:

0.001 mBar to 10 Bar ABS

House Material:

316 L Stainless Steel

Diameter:

25 mm

Length:

68.8 mm

Weight With Battery:

80 Grams

Memory Capacity:

60,000 Data Points

Minimum Sample Rate:

1 Second

Maximum Sample Rate:

24 Hours

Maximum Start Delay:

14 Days

Intrinsically Safe:

Ex II1GD Ex ia IIC T3 Ga,  -55°C ≤ Tamb ≤ +105°C

Time Accuracy:

± 5 Seconds Per 24 Hours

Battery:

TSP 150L Battery

If equipment is used in ATEX environment, special conditions for safe use stated in ATEX certificates section 17 must be considered.