TrackSense Pro XL Wireless Data Logger - Relative Humidity & Temperature Sensor

TrackSense Pro XL Wireless Data Logger - Relative Humidity & Temperature Sensor

The wireless TrackSense® Pro XL data loggers consist of resistant stainless steel and offers high-end technology, which provides highly accurate measurements within different thermal processes. The XL logger is a larger logger with an extended battery capacity, which makes it ideal for applications involving vast swings in temperature.

The Relative Humidity and Temperature Sensor (Stability) has a temperature range of 0 to +90 °C with an unmatched accuracy of ±0.1 °C (in range 0 to +85 °C) and a Relative humidity range of 0 to 100% RH. The sensor can be used for applications within the pharma industry, and is especially ideal to determine humidity levels during the validation of a warehouse or stability chambers.

Ellab’s TrackSense® Pro data loggers can all be configured with interchangeable sensors to measure pressure, relative humidity, temperature, CO2, and conductivity. The data loggers are easily activated and read by the Multi reader station and when combined with the Sky module they provide real time date via wireless communication.

 

Contact Sales
Technical Specifications

Sensor:

Interchangeable RH 5.0 Stability*

RH Measuring Range (0 to +85 °C):

0 - 100%

Measuring Principle:

Capacitance Sensing / Electrical Resistance

Sensor Element:

Capacitance / Pt1000 

RH Accuracy at 25°C non-condensing (10% to 90%):

± 2%

RH Resolution:

0.10 %

RH Response Time (63% in slowly moving air):

8 Sec.

Temperature Measuring Range:

0 to +90 °C

Temperature Accuracy:

+5 to +85 °C:

± 0.1 °C

Temperature Response Time:

<5 Sec.

Logger:

Pro XL

Operating Temperature:

0 to +90 °C

House Material:

316 L Stainless Steel

Diameter:

25 mm

Length:

68.8 mm

Weight With Battery:

80 Grams

Memory Capacity:

60,000 Data Points

Minimum Sample Rate:

1 Second

Maximum Sample Rate:

24 Hours

Maximum Start Delay:

14 Days

Intrinsically Safe:

Ex II1GD Ex ia IIC T3 Ga,  -55°C ≤ Tamb ≤ +105°C

Time Accuracy:

± 5 Seconds Per 24 Hours

Battery:

TSP 150L Battery

*With ValSuite versions prior to 5.X, these sensors will be listed as LAB sensors.