jpgFeature/sdesk 696 implement raw brick http
TrackSense Pro XL Wireless Data Logger - Thermocouple Sensor

TrackSense Pro XL Wireless Data Logger - Thermocouple Sensor

Ellab’s unique Thermocouple Sensor (TC Sensor) now allows users worldwide to use thermocouples in conjunction with wireless data loggers. The TC Sensor is especially designed for applications that cause a lot of strain on sensors and require as little impact on samples as possible. A good example of said applications would be Lyophilization, also known as Freeze Drying.

Replace Thermocouple Sensors with Ease
The feature that makes this sensor highly unique, is the ability to have the thermocouple wire replaced via a screw terminal after a prolonged series of use. Typically, if wires get strained through rough handling or immensely demanding applications, the entire sensor head would have to be replaced – now with the TC sensor, bundles of thermocouple wires can easily be acquired through Ellab and replace the single damaged wire, resulting in a far more cost-efficient process.

Thermocouple Wire
The thin thermocouple wire is equipped with a “naked tip” that takes up next to no room within a sample and provides users with a faster response time. The sensor itself has a temperature range of -80 to +62°C (calibrated from -60 to +60°C) with a state of the art accuracy of ±0.3°C. In addition to this, the wire comes in 10, 30, 50 and 100 cm lengths (30 cm standard), offering an additional layer of versatility.

Key Features & Benefits
✓  
Compatible with LYO fitting for Freeze Drying
✓  Ideal for applications that highly strain wires
✓  Can be cleaned by autoclave sterilization
✓  Replaceable thermocouple wires

 

Contact Sales

Technical Specifications

Sensor:

Thermocouple (TC) Sensor

Temperature Measuring Range:

-80 to +62 °C (Calibrated from -60 to +60°C)

Measuring Principle:

Thermoelectric Effect

Sensor Element:

Thermocouple

Accuracy:

± 0.3°C Calibrated

Position of Measuring Point from Tip:

0 mm

Cable Length:

10, 30, 50 or 100 cm

Cable dimensions and type:

0.55 x 0.95 mm type AWG30 with naked tip

Sensor Response Time:

T-63%:

0.03 Second

T-90%:

0.09 Second

Logger:

Pro XL

Operating Temperature:

-80 to +140 °C

Operating Pressure*:

0.0001 mBar to 4 Bar ABS

House Material:

316 L Stainless Steel

Diameter:

25 mm

Lenght:

68.8 mm

Weight With Battery:

80 Grams

Memory Capacity:

60,000 Data Points

Minimum Sample Rate:

1 Second

Maximum Sample Rate:

24 Hours

Maximum Start Delay:

14 Days

Intrinsically Safe:

Ex II1GD Ex ia IIC T3 Ga,  -50°C ≤ Tamb ≤ +105°C

Time Accuracy:

± 5 Seconds Per 24 Hours

Battery:

TSP 150L Battery

* Require ValSuite™ 5.1.0.11 or later, and logger firmware version 26.9 or later.